Function
Mode Range Accuracy Maximum permissible input
AC
current
A
RMS
Effective
value
30.00
±1.0%rdg.±0.7%f.s. 600A rms continuous
300.0
±1.0%rdg.±5dgt. 1000A max.
1000
±1.0%rdg.±5dgt.*1
(600)
(±1.0%rdg.±5dgt.*1)
1000A
rms 5 minutes
(600A
rms continuous 1000A max.)
PEAK
Peak
value
30.0
±5.0%rdg.±5dgt. 600A rms continuous
300
±3.0%rdg.±5dgt. 1000A max.
1000
±3.0%rdg.±5dgt.
(600)
(±3.0%rdg.±5dgt.)
1000A
rms 5 minutes
(600A
rms continuous 1000A max.)
AC
voltage
V
RMS 300.0/600
±1.0%rdg.±3dgt.*2
600V
rms continuous
PEAK
300/600
±3.0%rdg.±5dgt. 1000V max.
Crest
factor 1.00
to 5.00 ±10.0%rdg.±5dgt.
Effective
in the voltage
Frequency
Hz
and
current functions
100.0
±0.3%rdg.±1dgt.
1000
±1.0%rdg.±1dgt
Resistance
Ù 1k/10.00k
±1.5%rdg.±5dgt.
Overload
protection : 600V
rms
Open
terminal voltage:
3.0V DC max.
(resistance,continuity
functions)
Temperature
probe
sold separately
°C
-50
to 150°C ±2.0%rdg.±2dgt.
°F
-58
to 302°F Add the accuracy of the 9462
Continuity
1kÙ Buzzer sounds at approx 30Ù
or
less
*1.
40 to 45Hz , 66 to 1kHz ±1.5%rdg.±5dgt. *2. 40 to 45Hz , 66 to
1kHz ±1.5%rdg.±3dgt.
Display
LCD,